The most common complication of either type of jet ventilation is barotrauma, and the most common underlying etiology is unrecognized obstruction to outflow, either through the natural airways or in some circumstances through an endotracheal tube. This is counter to a common misconception that the high-pressure gas source itself is the cause of barotrauma. As discussed earlier, the high-pressure gas is essentially used as work to overcome the resistance of the small diameter ventilator delivery tubing and whatever jet device is attached to it. The pressures generated at the point of exit are significantly lower (consider that the hallmark of HFJV is
lower peak airway pressures than in conventional ventilation) than the set driving pressure. A frequently used example is that with a set driving pressure of 20 psi, a standard length of delivery tubing, and incorporating a 14G catheter, one could expect to deliver between 500 and 600 mL with a 1 second inspiration.
18,
19 Pressure in the lungs, then, depends on the volume delivered during each breath. The greater the volume delivered, the greater the pressure. A review of the literature reveals that the most common occurrence in cases of barotrauma is the development of an obstruction, often in the upper airway, which impedes egress.
20,
21,
22,
23,
24,
25,
26 If insufflation continues and this obstruction remains unrecognized, barotrauma ensues. This risk is higher in the setting of LFJV as the detection of impaired exhalation depends on the vigilance of the provider delivering the manual insufflations. HFJV offers the advantage of an integral alarm system designed to detect outflow obstruction. Modern high-frequency jet ventilators incorporate a sophisticated switching system that enables the delivery tubing itself to act as pressure tubing to a dedicated pressure transducer. At the end of the expiratory cycle, backpressure in the airway is measured, and if the set alarm limit is exceeded, delivery of additional breaths is stopped. This occurs at the end of each respiratory cycle, regardless of the set rate, providing breath-to-breath detection of potential outflow obstruction. The alarm level is adjustable but is often nondefeatable. A typical setting for this alarm limit is 20 cm H
2O.