The myriad electronic devices in the modern operating room (OR) greatly improve patient care and safety but also subject patients and OR personnel to increased risks (Ehrenwerth J, Seifert HA. Electrical and fire safety. In: Barash PG, Cullen BF, Stoelting RK, Cahalan MK, Ortega R, Stock MC, eds. Clinical Anesthesia. Philadelphia: Lippincott Williams & Wilkins; 2013: 189–218).


I. Principles of Electricity

A basic principle of electricity is known as Ohm’s law and is represented by the equation E = I × R (electromotive force in volts = current in amperes times resistance in ohms). Ohm’s law forms the basis for the physiologic equation in which the blood pressure is equal to the cardiac output times the systemic vascular resistance (BP = CO × SVR). Electrical power is measured as watts (voltage × amperage). The amount of electrical work done (watt-second or joule) is a common designation for electrical energy expended. (Energy produced by a defibrillator is measured in joules.)



  • Direct and Alternating Currents. The flow of electrons (current) through a conductor is characterized as direct current (electron flow is always in the same direction) or alternating current (electron flow reverses direction at a regular interval).


  • Impedance is the sum of forces that oppose electron movement in an alternating current circuit.


  • Capacitance is the ability of a capacitor (two parallel conductors separated by an insulator) to store charge.



    • In a direct current circuit, the charged capacitor plates (battery) do not result in current flow unless a resistance is connected between the two plates and the capacitor is discharged.


    • Stray capacitance is capacitance that is not designed into the system but is incidental to the construction of the equipment. All alternating current operating equipment produces stray capacitance even while turned off.


II. Electrical Shock Hazards



  • Alternating and Direct Currents



    • Whenever an individual contacts an external source of electricity, an electrical shock is possible. (It requires
      approximately three times as much direct current as alternating current to cause ventricular fibrillation [VF].)


    • A short circuit occurs when there is zero impedance with a high current flow.


  • Source of Shocks



    • Electrical accidents or shocks occur when a person becomes part of or completes an electrical circuit (Fig. 8-1).


    • Damage from electrical current is caused by disruption of normal electrical function of cells (skeletal muscle contracture, VF) or dissipation of electrical energy (burn).


    • The severity of an electrical shock is determined by the amount of current and the duration of current flow.



      • Macroshock describes large amounts of current flow that can cause harm or death.


      • Microshock describes small amounts of current flow and applies only to electrically susceptible patients (those with an external conduit that is in direct contact with the heart, such as a pacing wire or saline-filled central venous pressure [CVP] catheter) in whom even minute amounts of current (1 mA, which is the threshold of perception) may cause VF.


    • Very high-frequency current does not excite contractile tissue and does not cause cardiac dysrhythmias.



  • Grounding. To fully understand electrical shock hazards and their prevention, one must have a thorough knowledge of the concepts of grounding. In electrical terminology, grounding is applied to electrical power and equipment.






Figure 8-1. When a faulty piece of equipment without an equipment ground wire is plugged into an electrical outlet not containing a ground wire, the instrument case becomes energized (“hot”). If an individual touches the case (A), he or she will receive a shock (dashed line depicts path of electrical current) because he or she is standing on the ground (B) and completes the circuit.


III. Electrical Power: Grounded



  • Electrical utilities universally provide power to homes that are grounded. (By convention, the earth ground potential is zero.)


  • Electrical shock is an inherent danger of grounded power systems. An individual standing on ground or in contact with an object that is referenced to the ground needs only one additional contact point to complete the circuit.


  • Modern wiring systems have added a third wire (a low-resistance pathway through which the current can flow to ground) to decrease the severity of potential electrical shocks (Fig. 8-2).


IV. Electrical Power: Ungrounded



  • The numerous electronic devices, along with power cords and puddles of saline-filled solutions on the floor, tend to make
    the OR an electrically hazardous environment for both patients and personnel.


  • In an attempt to decrease the risk of electrical shock, the power supplied to most ORs is ungrounded (current is isolated from the ground).


  • Supplying ungrounded power to an OR requires the use of an isolation transformer (Fig. 8-3).



    • The isolated power system provides protection from macroshock (Fig. 8-4).



    • A faulty piece of equipment plugged into an isolated power system does not present a shock hazard.






Figure 8-2. When a faulty piece of equipment containing an equipment ground wire is properly connected to an electrical outlet with grounding protection, the electrical current (dashed line) will preferentially flow down the low-resistance ground wire. An individual touching the instrument case (A) and standing on the ground (B) still completes the circuit; however, only a small part of the current flows through the individual.






Figure 8-3. In an operating room, the isolation transformer converts the grounded power on the primary side to an ungrounded power system on the secondary side of the transformer. There is no direct connection from the power on the secondary side to ground. The equipment ground wire, however, is still present.






Figure 8-4. A safety feature of the isolated power system is illustrated. An individual contacting one side of the isolated power system (A) and standing on the ground (B) will not receive a shock. In this instance, the individual is not contacting the circuit at two points and thus is not completing the circuit.


V. The Line Isolation Monitor