CHAPTER 27 Arterial Catheterization and Pressure Monitoring





2 How do noninvasive blood pressure devices work?


Blood pressure is usually measured either manually (ausculatory method) or with an automated device (oscillometric method). With the ausculatory method, a pneumatic cuff is inflated to occlude arterial blood flow. As the cuff is deflated, audible frequencies called Korotkoff sounds are created by turbulent blood flow in the artery. The pressure at which the sounds are first audible is taken as the systolic pressure, and the pressure at which the sounds become muffled or disappear is taken as the diastolic pressure. Errors in measurement may be caused by:









With the oscillometric method, a pneumatic cuff is also inflated to occlude the arterial blood flow. As the cuff is deflated, the arterial pulsations cause pressure changes in the cuff that are analyzed by a computer. The systolic pressure is taken as the point of rapidly increasing oscillations, the mean arterial pressure as the point of maximal oscillation, and the diastolic pressure as the point of rapidly decreasing oscillations. Errors in measurement may occur from inappropriate cuff size or factors that prevent detection of cuff pressure variations, such as patient shivering. Prolonged use of the stat mode, in which the cuff reinflates immediately after each measurement is obtained, may lead to complications such as ulnar nerve paresthesia, thrombophlebitis, or compartment syndrome.





5 How is radial artery catheterization performed?


The wrist is dorsiflexed and immobilized, the skin is cleaned with an antiseptic solution, the course of the radial artery is determined by palpation, and local anesthetic is infiltrated into the skin overlying the artery (if the patient is awake). A 20-G over-the-needle catheter apparatus is inserted at a 30- to 45-degree angle to the skin along the course of the radial artery. After arterial blood return, the angle is decreased, and the catheter is advanced slightly to ensure that both the catheter tip and the needle have advanced into the arterial lumen. The catheter is then threaded into the artery. Alternatively the radial artery may be transfixed. After arterial blood return, the apparatus is advanced until both the catheter and the needle pass completely through the front and back walls of the artery. The needle is withdrawn into the catheter, and the catheter is pulled back slowly. When pulsatile blood flow is seen in the catheter, it is advanced into the lumen. If the catheter will not advance into the arterial lumen and blood return is good, a sterile guidewire may be placed into the lumen through the catheter, and the catheter advanced over the wire.

May 31, 2016 | Posted by in ANESTHESIA | Comments Off on 27: Arterial Catheterization and Pressure Monitoring

Full access? Get Clinical Tree

Get Clinical Tree app for offline access