19: Anesthesia Circuits and Ventilators

CHAPTER 19 Anesthesia Circuits and Ventilators






3 Give an example of a semiopen circuit


The various semiopen circuits were fully described by Mapleson and are commonly known as the Mapleson A, B, C, D, E, and F circuits (Figure 19-1). All have in common a source of fresh gas, corrugated tubing (more resistant to kinking), and a pop-off or adjustable pressure-limiting valve. Differences among the circuits include the locations of the pop-off valve, fresh gas input, and whether or not a gas reservoir bag is present. Advantages of the Mapleson series are simplicity of design, ability to change the depth of anesthesia rapidly, portability, and lack of rebreathing of exhaled gases (provided the fresh gas flow is adequate). Disadvantages include lack of conservation of heat and moisture, limited ability to scavenge waste gases, and high requirements for fresh gas flow. These are rarely used today except for patient transport.




4 Give an example of a semiclosed circuit


The prototypical semiclosed circuit is the circle system, which is found in most operating rooms in the United States (Figure 19-2). Every semiclosed system contains an inspiratory limb, expiratory limb, unidirectional valves, carbon dioxide (CO2) absorber, gas reservoir bag, and a pop-off valve on the expiratory limb. Advantages of a circle system include conservation of heat and moisture, the ability to use low flows of fresh gas (thereby conserving volatile anesthetic and the ozone layer), and the ability to scavenge waste gases. A disadvantage is its complex design; it has approximately 10 connections, each of which has the potential for disconnection.






May 31, 2016 | Posted by in ANESTHESIA | Comments Off on 19: Anesthesia Circuits and Ventilators

Full access? Get Clinical Tree

Get Clinical Tree app for offline access